Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes

Received: 15 April 2021    Accepted: 30 April 2021    Published: 14 May 2021
Views:       Downloads:
Abstract

Many diabetic mellitus patients envision a non-invasive method of blood glucose measurement since they require periodic monitoring of their blood glucose levels to ensure that it is stable and within the normal range. In this study, we detected glucose concentration using commercial light emitting diodes (LEDs) with a wavelength of 700–1600 nm. Light of long wavelengths (e.g. NIR) infiltrate human skin and reach the blood vessel lining, thereby aiding in the non-invasive measurement of blood glucose concentration. To demonstrate this, the concentration of glucose solutions was measured using red and NIR-LEDs in a non-invasive manner. The sensitivity of glucose detection was greater when light of wavelength below 1000 nm was used, owing to the absorption of wavelengths above 1000 nm by water. Furthermore, we controlled the input current of the red and NIR-LEDs to confirm the light intensity deviation with increasing glucose concentrations and suggested the optimum wavelength of light using this in-vitro system. Among various LEDs, the 700 nm LED showed higher light intensity deviation with change in injection current compared to LEDs with the other wavelengths. In particular, compared to other LEDs, a stark difference was observed in the light intensity of the 700 nm LED while measuring glucose concentrations in the range of 50–100 mg/dl.

Published in Biomedical Sciences (Volume 7, Issue 2)
DOI 10.11648/j.bs.20210702.12
Page(s) 47-52
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2024. Published by Science Publishing Group

Keywords

Non-Invasive, Glucose, NIR-LED, Diabetic Mellitus, Water Absorption

References
[1] http://www.diabetesatlas.org/ (2015).
[2] Ong, W. M., Chua, S. S., Ng, C. J. (2014). Barriers and facilitators to self-monitoring of blood glucose in people with type 2 diabetes using insulin: a qualitative study. Patient Prefer. Adherence, 8, 237-246.
[3] http://www.diabetes.co.uk/diabetes_care/blood-sugar-level-ranges.html (2019).
[4] Kennedy, L. (2001). Self-Monitoring of Blood Glucose in Type 2 Diabetes Time for evidence of efficacy. Diabetes Care, 24 (6): 977–978.
[5] American Diabetes Association, (2015). Standards of Medical Care in Diabetes—2015 Abridged for Primary Care Providers. Clinical Diabetes, 33 (2): 97-111.
[6] http://www.diabetesnet.com/diabetes-technology/metesr-monitors/continuous-monitors/compare-current-monitors (2018).
[7] McMillin, J. M. (1990). Clinical methods: The history, physical, and laboratory examinations Blood Glucose, 3rd ed., Boston, MA, USA: Butterworth, ch. 141, 662-665.
[8] Wang, H. -Ch., Lee, A. R. (2015). Recent developments in blood glucose sensors. J. Food Drug Anal., 23 (2): 191–200.
[9] Klonoff, D. C. (2007). The Benefits of Implanted Glucose Sensors. J. Diabetes Sci. Technol., 6, 797-800.
[10] Mamkin, I., Ten, S., Bhandari, S., Ramchandani, N. (2008). Real-time continuous glucose monitoring in the clinical setting: the good, the bad, and the practical. J. Diabetes Sci. Technol., 2 (5): 882-889.
[11] So, C.-F., Choi, K.-S., Wong, T. K. S., Chung, J. W. Y. (2012). Recent advances in noninvasive glucose monitoring. Med. Devices Evidence Res., 5, 45-52.
[12] Tura, A., Sbrignadello, S., Clanciavicchia, D., Pacini, G., Ravazzani, P. (2010). A Low Frequency Electromagnetic Sensor for Indirect Measurement of Glucose Concentration: In Vitro Experiments in Different Conductive Solutions. Sensors, 10 (6): 5346-5358.
[13] Yatim, N. N. M., Yatim, M., Zain, Z. M., Jaafar, M. Z., Yusof, Z. Md. (2014). Noninvasive glucose level determination using diffuse reflectance near infrared spectroscopy and chemometrics analysis based on in vitro sample and human skin. In Proc. IEEE Conf. Syst., Process Control, Kuala Lumpur, Malaysia, 30-35.
[14] Shao J., Lin, M., Li, Y., Li, X., Liu, J., Liang, J., Yao, H. (2012). In Vivo Blood Glucose Quantification Using Raman Spectroscopy, PLoS One, 7 (10): e48127.
[15] Helse H. M., Winzen, R., Fundamental chemometric methods. In Near Infrared Spectroscopy: Principles, Instruments, Applications, (2002). H. W. Siesler, Y. Ozaki, S. Kawata, and H. M. Heise, Eds. Weinheim, Germany: Willey, ch. 7, 125-162. Yung, R., & Khoo-Lattimore, C. (2019). New realities: a systematic literature review on virtual reality and augmented reality in tourism research. Current Issues in Tourism, 22, 2056 - 2081.
[16] T. Kossowski, K. Kossowska, R. Stasiński, (2017). Non-invasive measurement of chemical compounds levels in human body. Measurement Automation Monitoring, 63 (06): 200-202.
[17] S. Robiah, R. P. Jenie, M. Dahrul, N. M. Nurdin, J. Iskandar, A. Kurniawan, E. Rustami, H. Syafutra, H. Alatas, Irzaman, E. Damayanthi, Rimbawan, D. Sukandar, Evriyanti, and S. Budiarti, (2017). Infra Red Light Emitting Diode in 1200 nm Range have Moderate Performance in Detecting Glucose in Human Blood Glucose Model, IOP Conf. Series: Earth and Environmental Science, 58 (012021): 1-4.
[18] American Diabetes Association. Standards of Medical Care in Diabetes (2015). Diabetes Care, 38 (Suppl. 1), S1-S94.
[19] R. Hotmartua, P. W. Pangestu, H. Zakaria, Y. S. Irawan, (2015). Noninvasive blood glucose detection using near infrared sensor. The 5th International Conference on Electrical Engineering and Informatics, IEEE, 978, 687-692.
[20] Sekar, S. K. V., Bargigia, I., Mora, A. D., Taroni, P., Ruggeri, A., Tosi, A., Pifferi, A., Farina, A., (2017). Diffuse optical characterization of collagen absorption from 500 to 1700 nm, J. Biomed. Opt., 22 (1): 015006.
[21] S. K. Vashist, (2012). Non-invasive glucose monitoring technology in diabetes management: a review, Anal. Chim. Acta, 750, 16-27.
[22] S. Perrey, (2008). Non-invasive NIR spectroscopy of human brain function during exercise, Methods, 45, 289-299.
[23] Zirk, K., Poetzschke, H. (2004). On the suitability of refractometry for the analysis of glucose in blood-derived fluids. Med. Eng. Phys., 26, 473-481.
[24] O. S. Abdalsalam, A. A. Awouda, (2014). Noninvasive glucose monitoring using scattering spectroscopy. Am. J. Biomed. Eng., 4 (3): 53-59.
[25] Haxha, S., Jhoja J. (2016). Optical Based Noninvasive Glucose Monitoring Sensor Prototype. IEEE Photonics J., 8 (6): 6805911.
[26] McEwen, M. P., Reynolds, K. J. (2014). Noninvasive Monitoring With Strongly Absorbed Light. OPT. APPL., XLIV, 2, 177-190.
[27] Khalil, O. S. (1999). Spectroscopic and clinical aspects of noninvasive glucose measurements. Clin. Chem., 45 (2): 165-177.
Cite This Article
  • APA Style

    Semi Oh, Min-Seok Kim, Beom-Rae Noh, Hyung-Hwan Baik, Kyoung-Kook Kim. (2021). Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes. Biomedical Sciences, 7(2), 47-52. https://doi.org/10.11648/j.bs.20210702.12

    Copy | Download

    ACS Style

    Semi Oh; Min-Seok Kim; Beom-Rae Noh; Hyung-Hwan Baik; Kyoung-Kook Kim. Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes. Biomed. Sci. 2021, 7(2), 47-52. doi: 10.11648/j.bs.20210702.12

    Copy | Download

    AMA Style

    Semi Oh, Min-Seok Kim, Beom-Rae Noh, Hyung-Hwan Baik, Kyoung-Kook Kim. Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes. Biomed Sci. 2021;7(2):47-52. doi: 10.11648/j.bs.20210702.12

    Copy | Download

  • @article{10.11648/j.bs.20210702.12,
      author = {Semi Oh and Min-Seok Kim and Beom-Rae Noh and Hyung-Hwan Baik and Kyoung-Kook Kim},
      title = {Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes},
      journal = {Biomedical Sciences},
      volume = {7},
      number = {2},
      pages = {47-52},
      doi = {10.11648/j.bs.20210702.12},
      url = {https://doi.org/10.11648/j.bs.20210702.12},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.bs.20210702.12},
      abstract = {Many diabetic mellitus patients envision a non-invasive method of blood glucose measurement since they require periodic monitoring of their blood glucose levels to ensure that it is stable and within the normal range. In this study, we detected glucose concentration using commercial light emitting diodes (LEDs) with a wavelength of 700–1600 nm. Light of long wavelengths (e.g. NIR) infiltrate human skin and reach the blood vessel lining, thereby aiding in the non-invasive measurement of blood glucose concentration. To demonstrate this, the concentration of glucose solutions was measured using red and NIR-LEDs in a non-invasive manner. The sensitivity of glucose detection was greater when light of wavelength below 1000 nm was used, owing to the absorption of wavelengths above 1000 nm by water. Furthermore, we controlled the input current of the red and NIR-LEDs to confirm the light intensity deviation with increasing glucose concentrations and suggested the optimum wavelength of light using this in-vitro system. Among various LEDs, the 700 nm LED showed higher light intensity deviation with change in injection current compared to LEDs with the other wavelengths. In particular, compared to other LEDs, a stark difference was observed in the light intensity of the 700 nm LED while measuring glucose concentrations in the range of 50–100 mg/dl.},
     year = {2021}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Non-Invasive Measurement of Glucose Concentration Using Red and Near-Infrared Light-Emitting Diodes
    AU  - Semi Oh
    AU  - Min-Seok Kim
    AU  - Beom-Rae Noh
    AU  - Hyung-Hwan Baik
    AU  - Kyoung-Kook Kim
    Y1  - 2021/05/14
    PY  - 2021
    N1  - https://doi.org/10.11648/j.bs.20210702.12
    DO  - 10.11648/j.bs.20210702.12
    T2  - Biomedical Sciences
    JF  - Biomedical Sciences
    JO  - Biomedical Sciences
    SP  - 47
    EP  - 52
    PB  - Science Publishing Group
    SN  - 2575-3932
    UR  - https://doi.org/10.11648/j.bs.20210702.12
    AB  - Many diabetic mellitus patients envision a non-invasive method of blood glucose measurement since they require periodic monitoring of their blood glucose levels to ensure that it is stable and within the normal range. In this study, we detected glucose concentration using commercial light emitting diodes (LEDs) with a wavelength of 700–1600 nm. Light of long wavelengths (e.g. NIR) infiltrate human skin and reach the blood vessel lining, thereby aiding in the non-invasive measurement of blood glucose concentration. To demonstrate this, the concentration of glucose solutions was measured using red and NIR-LEDs in a non-invasive manner. The sensitivity of glucose detection was greater when light of wavelength below 1000 nm was used, owing to the absorption of wavelengths above 1000 nm by water. Furthermore, we controlled the input current of the red and NIR-LEDs to confirm the light intensity deviation with increasing glucose concentrations and suggested the optimum wavelength of light using this in-vitro system. Among various LEDs, the 700 nm LED showed higher light intensity deviation with change in injection current compared to LEDs with the other wavelengths. In particular, compared to other LEDs, a stark difference was observed in the light intensity of the 700 nm LED while measuring glucose concentrations in the range of 50–100 mg/dl.
    VL  - 7
    IS  - 2
    ER  - 

    Copy | Download

Author Information
  • Department of Medicine, Kyung Hee University, Seoul, Republic of Korea

  • Department of Medicine, Kyung Hee University, Seoul, Republic of Korea

  • Department of Nano & Semiconductor Engineering, Korea Polytechnic University, Siheung, Republic of Korea

  • Department of Medicine, Kyung Hee University, Seoul, Republic of Korea

  • Department of Nano & Semiconductor Engineering, Korea Polytechnic University, Siheung, Republic of Korea

  • Sections