| Peer-Reviewed

Assessment of Frequency of Errors in Conventional Panoramic Radiographs

Received: 25 October 2016    Accepted: 8 November 2016    Published: 10 December 2016
Views:       Downloads:
Abstract

Although use of conventional panoramic systems leads to degrade the image quality, it is still common in Sri Lanka. The objective of the study was to determine the frequency of errors occurred during Ortho Pantomo Graph (OPG) and to minimize the further occurrence. Two hundred and fifty OPGs were collected from Department of Radiology, Dental Hospital (Teaching) Peradeniya (DHTP). Quality of 250 OPGs was evaluated by one radiologist and a radiographer. Evaluators assessed all 250 radiographs for presence of errors in categories including patient preparation, patient positioning, patient motion, exposure, handling and processing. The frequency of each error was recorded. Further, number of repeated radiographs was also recorded. There were only 22 (9%) error-free OPGs. 228 (91%) radiographs had errors. 6% of OPGs were unacceptable and had to be re-taken. The frequencies of errors in conventional OPGs were relatively high in this study with patient positioning errors being the most frequent error.

Published in International Journal of Dental Medicine (Volume 2, Issue 2)
DOI 10.11648/j.ijdm.20160202.11
Page(s) 5-8
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2024. Published by Science Publishing Group

Keywords

Radiography, Ortho Pantomo Graph, Quality Control

References
[1] Choi BR, Choi DH, Huh KH, Yi WJ, Heo MS, Choi SC, et al. Clinical image quality evaluation for panoramic radiography in Korean dental clinics. Imaging Sci Dent. 2012; 42 (3): 183–90.
[2] Whitley, A. Stewart; Sloane, Charles; Hoadley, Graham; Moore, Adrian D.; Alsop CW. Clark’s Positioning in Radiography. Vasa. 2005; 126.
[3] Eric whites, Churchill Livingstone Elsevier, Essentials of dental radiography and radiology, dental panoramic radiography (4th edn), Philadelphia, USA, 2007.
[4] Haring JI, Howerton LJ, Panoramic radiography, Dental radiography principles and techniques (3rd edn), Elsevier, India, P A: Saunders, 2006, p p 304-322.
[5] Jayasinghe R. D., Oral radiography and radiology for dental students (2012), dental panoramic radiography (1st edn), Sri Lanka.
[6] Langland O. E, Sippy F. H., Morris C. R., and Langlais R. P., Principles and Practice of Panoramic Radiology, W B Saunders, London, UK, 2nd edition, 1992.
[7] Peretz B, Gotler M. and Kaffe I, Common Errors in Digital Panoramic Radiographs of Patients with Mixed Dentition and Patients with Permanent Dentition, International Journal of Dentistry, Volume 2012 (2012), Article ID 584138, doi:10.1155/2012/584138 Research Article.
[8] William C. Scarfe, Gail F. Williamson, Practical Panoramic Radiography, Revised July 25, 2007, Available: http://media.dentalcare.com/media/enus/education/ce71/ce71.pdf
[9] Dhillon M, Raju SM, Verma S, Tomar D, Mohan RS, Lakhanpal M, et al. Positioning errors and quality assessment in panoramic radiography. Imaging Sci Dent. 2012;42 (4): 207–12.
[10] Manufacture’s guidelines, Planmeca Proline XC radiographic machine, asentajankatu6, 00880 Helsinki, Finland.
[11] Sabarudin A, Tiau YJ. Image quality assessment in panoramic dental radiography: a comparative study between conventional and digital systems. Quant Imaging Med Surg. 2013; 3(1): 43–8.
[12] Brezden NA, Brooks SL. Evaluation of panoramic dental radiographs taken in private practice. Oral Surg Oral Med Oral Pathol. 1987; 63: 617–621.
[13] Langland OE, Langlais RP, Preece JW. Principles of Dental Imaging, 2nd ed. Philadelphia: Lippincott Williams & Wilkins; 2002: 219-258.
[14] Kaviani F, Johari M, Esmaeili F. Evaluation of common errors of panoramic radiographs in tabriz faculty of dentistry. J Dent Res Dent Clin Dent Prospects [Internet]. 2008; 2 (3): 99–101.
Cite This Article
  • APA Style

    Abedeera Jayasuriya Seena Patabedige Nileema, Thasanthan Loganathan, Dissanayake Mudiyanselage Janaka Nuwan. (2016). Assessment of Frequency of Errors in Conventional Panoramic Radiographs. International Journal of Dental Medicine, 2(2), 5-8. https://doi.org/10.11648/j.ijdm.20160202.11

    Copy | Download

    ACS Style

    Abedeera Jayasuriya Seena Patabedige Nileema; Thasanthan Loganathan; Dissanayake Mudiyanselage Janaka Nuwan. Assessment of Frequency of Errors in Conventional Panoramic Radiographs. Int. J. Dent. Med. 2016, 2(2), 5-8. doi: 10.11648/j.ijdm.20160202.11

    Copy | Download

    AMA Style

    Abedeera Jayasuriya Seena Patabedige Nileema, Thasanthan Loganathan, Dissanayake Mudiyanselage Janaka Nuwan. Assessment of Frequency of Errors in Conventional Panoramic Radiographs. Int J Dent Med. 2016;2(2):5-8. doi: 10.11648/j.ijdm.20160202.11

    Copy | Download

  • @article{10.11648/j.ijdm.20160202.11,
      author = {Abedeera Jayasuriya Seena Patabedige Nileema and Thasanthan Loganathan and Dissanayake Mudiyanselage Janaka Nuwan},
      title = {Assessment of Frequency of Errors in Conventional Panoramic Radiographs},
      journal = {International Journal of Dental Medicine},
      volume = {2},
      number = {2},
      pages = {5-8},
      doi = {10.11648/j.ijdm.20160202.11},
      url = {https://doi.org/10.11648/j.ijdm.20160202.11},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.ijdm.20160202.11},
      abstract = {Although use of conventional panoramic systems leads to degrade the image quality, it is still common in Sri Lanka. The objective of the study was to determine the frequency of errors occurred during Ortho Pantomo Graph (OPG) and to minimize the further occurrence. Two hundred and fifty OPGs were collected from Department of Radiology, Dental Hospital (Teaching) Peradeniya (DHTP). Quality of 250 OPGs was evaluated by one radiologist and a radiographer. Evaluators assessed all 250 radiographs for presence of errors in categories including patient preparation, patient positioning, patient motion, exposure, handling and processing. The frequency of each error was recorded. Further, number of repeated radiographs was also recorded. There were only 22 (9%) error-free OPGs. 228 (91%) radiographs had errors. 6% of OPGs were unacceptable and had to be re-taken. The frequencies of errors in conventional OPGs were relatively high in this study with patient positioning errors being the most frequent error.},
     year = {2016}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Assessment of Frequency of Errors in Conventional Panoramic Radiographs
    AU  - Abedeera Jayasuriya Seena Patabedige Nileema
    AU  - Thasanthan Loganathan
    AU  - Dissanayake Mudiyanselage Janaka Nuwan
    Y1  - 2016/12/10
    PY  - 2016
    N1  - https://doi.org/10.11648/j.ijdm.20160202.11
    DO  - 10.11648/j.ijdm.20160202.11
    T2  - International Journal of Dental Medicine
    JF  - International Journal of Dental Medicine
    JO  - International Journal of Dental Medicine
    SP  - 5
    EP  - 8
    PB  - Science Publishing Group
    SN  - 2472-1387
    UR  - https://doi.org/10.11648/j.ijdm.20160202.11
    AB  - Although use of conventional panoramic systems leads to degrade the image quality, it is still common in Sri Lanka. The objective of the study was to determine the frequency of errors occurred during Ortho Pantomo Graph (OPG) and to minimize the further occurrence. Two hundred and fifty OPGs were collected from Department of Radiology, Dental Hospital (Teaching) Peradeniya (DHTP). Quality of 250 OPGs was evaluated by one radiologist and a radiographer. Evaluators assessed all 250 radiographs for presence of errors in categories including patient preparation, patient positioning, patient motion, exposure, handling and processing. The frequency of each error was recorded. Further, number of repeated radiographs was also recorded. There were only 22 (9%) error-free OPGs. 228 (91%) radiographs had errors. 6% of OPGs were unacceptable and had to be re-taken. The frequencies of errors in conventional OPGs were relatively high in this study with patient positioning errors being the most frequent error.
    VL  - 2
    IS  - 2
    ER  - 

    Copy | Download

Author Information
  • Department of Radiography/Radiotherapy, Faculty of Allied Health Sciences, University of Peradeniya, Peradeniya, Sri Lanka

  • Department of Radiography/Radiotherapy, Faculty of Allied Health Sciences, University of Peradeniya, Peradeniya, Sri Lanka

  • Department of Radiography/Radiotherapy, Faculty of Allied Health Sciences, University of Peradeniya, Peradeniya, Sri Lanka

  • Sections